CO₂, temperature & humidity

Three air measurements. One LoRaWAN sensor.

Request a quotation
CO₂, temperature & humidity
SenseCAP S2103 · Battery powered · IP66

Follow air conditions across your site

Monitor CO₂, temperature and humidity in production spaces, offices and storage areas. Battery power makes it easy to add measurement points. The M4 keeps their history locally, ready to review in your browser or connect to your building systems.

CO₂ concentration
400–10,000 ppm
Measurements
CO₂ + temperature + relative humidity
Installation
Battery powered · LoRaWAN · IP66

Three readings. A clearer view of the space.

CO₂

Ventilation patterns

Follow how CO₂ concentration rises during use and falls as the space clears or ventilation changes.

°C

Temperature changes

Compare air temperature between spaces and shifts, alongside the activity in your building.

% RH

Humidity trends

See when moisture levels change around production, stored materials or occupied rooms.

Choose a representative measurement point

Mount the probe facing down in representative airflow, away from direct sunlight and nearby heat. Keep the complete CO₂ sensor within 0–50 °C and non-condensing conditions.

See the working day in your measurements

Read all three values together, then open the history to compare changes between shifts or spaces. The example shows repeated CO₂ rises and recoveries across three days.

The sensor reports over LoRaWAN to a compatible interface on the M4. Your readings stay in its local database, with browser access and OPC UA or MQTT integration for your own systems.

Actual iQunet software · Illustrative readings at ten-minute intervals.

CO₂, temperature & humidity monitoring

Actual iQunet software · Illustrative readings at ten-minute intervals.

Swipe sideways for detail. Tap the image to close.

Technical details

Measurement specs

Product documents

Datasheet PDF

Request a quotation

CO₂, temperature & humidity

Tell us which spaces you want to monitor and how many measurement points you need. We’ll specify the sensors and matching LoRaWAN connection.

CO₂, temperature & humidity

Price on requestAvailability confirmed with your quotation

CO₂, temperature & humidity

Technical details

Measurement specifications

SenseCAP S2103 measurement and installation specifications

CO₂ concentration
400–10,000 ppm; 1 ppm resolution. Accuracy from 400 to 5,000 ppm: ±(30 ppm + 3% of reading). From 5,000 to 10,000 ppm: ±10% of reading.
Operating conditions
0 to +50 °C for CO₂ operation; 0–90% RH, non-condensing. IP66 enclosure.
Air temperature
Measurement channel: −40 to +85 °C; stated accuracy ±0.2 °C; 0.01 °C resolution. Keep the complete CO₂ sensor within its 0 to +50 °C operating range.
Relative humidity
Measurement channel: 0–100% RH; stated accuracy ±1.8% RH; 0.01% RH resolution. Observe the non-condensing operating limit.
Radio
LoRaWAN 1.0.3 Class A. Configurable EU868, US915, AU915, AS923, IN865, KR920 and RU864 plans. Match the sensor, receiving interface and installation country.
Battery
Replaceable 19 Ah, non-rechargeable D-size lithium thionyl chloride battery. Service life depends on the reporting interval, radio settings and conditions.
Size & mass
63 × 57 mm housing; 144.5 mm body plus 39.7 mm probe. Device mass: 285 g.
Setup & calibration
Bluetooth app configuration. CO₂ single-point calibration against a known reference concentration; follow the manufacturer procedure and maintenance guidance.

Connection & mounting

Private LoRaWAN and local data storage

Connect the supported sensor through the matching LoRaWAN interface. The edge server stores its decoded measurements locally and makes them available through OPC UA, MQTT and GraphQL.

SenseCAP S2103. Mount on a wall or pillar with the probe facing down and exposed to representative airflow. Avoid direct sunlight and nearby heat sources. Configure the reporting interval and matching radio region through the sensor's Bluetooth app, then commission it with the iQunet LoRaWAN interface.

Compatible components

Datasheets & installation

PDFS2103 specifications · manufacturer catalogue, page 7PDFSenseCAP S210X setup, mounting & calibration guide
iQunet · Engineering enquiries

Ask us a question

Ask about our products or pricing, request a demonstration, or get technical help.

About your question

About: CO₂, temperature & humidity

Request type: General question
Prefer email or a call?info@iqunet.com+32 52 86 00 25

We use your details to answer this request. Privacy information (new tab)